Quality Control
Each batch should include several analytical controls:
- Blanks to monitor chemical/lipid background and carryover as a sources of artefacts:
- Solvent blank
- Internal standard blank
- Extraction blank
- Quality control samples (QCs) to demonstrate comparable concentrations and coverage of lipid species between different batches within one or between different laboratories
- QCs and reference material should have the same matrix as the sample material
- Reference materials like NIST SRM 1950 are necessary to permit community wide comparability